Note:
1. The indicated standard values are a recommendation, the dentist determines the individual setting
of the exposure values. Different X-ray machine brands vary in structural design, and multiple factors
affect radiation dose. Actual exposure time settings should be determined through practical testing.
2. The above table shows the standard values for the exposure time of a PSP for an adult patient. For
patients with non-standard body habitus (underweight or overweight), moderate dose adjustments
should be applied accordingly.
3. Children are more sensitive to radiation than adults. Keep the exposure parameters as low as
possible taking into consideration the image quality. For pediatric patients, we recommend reducing
the exposure time by approximately 0.05 seconds compared to adult settings.
4. When all other parameters remain constant, the X-ray tube length also affects radiation dose. If the
tube length is more than 14cm, the dose should be increased accordingly. Specific adjustments should
be determined through actual measurement.